6. Specifications
No. 99MBA091A
6-7
6.7
Serial Data Output Specifications of the Counter
The measured date of the counter unit is output via the serial output connector to a peripheral,
such as two-dimensional data processing unit Micropak 9/7 or two-dimensional data
processing program QSPAK.
Communication Specifications
・
Transmission method
:
Half duplex
・
Transmission control
:
Asynchronous (Start-stop transmission)
・
Baud rate
:
1200bps
/
2400bps
/
4800bps
/
9600bps
(
※
1)
/
19200bps
・
Data bit
:
7 bits
(
※
1)
/
8 bits
ASCII
・
Parity bit
:
Even
(
※
1)
/
Odd
/
None
・
Stop bit
:
1 bit
(
※
1
)
/
2 bits
・
Flow control
:
Enabled
/
Disabled
(
※
1
)
(※1)
Default setting at the shipment
6.7.1 Connector
specifications
Connector shape
Pin
number
Signal
I/O
direction
※
2
Description
1
2 RD IN
Accept
the
command
3 TD OUT
Measured
data
4
DTR
OUT
Counter is ready
5 SG
-
Signal ground
6 DSR IN
Peripheral
is
ready
7
-
-
8
-
-
9
FG
-
-
Frame ground
(※2)
In/Out direction OUT
:
counter
→
Peripheral
The input/output circuits used are equivalent to MAX232 (MAXIM Corporation).
6.7.2 Control
signal
A data request from a peripheral is accepted only if the DTR signal is “H” (space).
“H” has a +9V level.
The counter unit can transfer only if the DSR signal is “H” (space).
D-sub, 9 pins,
male, inch thread
Summary of Contents for MF 1010C Series
Page 2: ......
Page 54: ...No 99MBA091A 3 14 MEMO ...
Page 66: ...No 99MBA091A 5 6 MEMO ...
Page 82: ...No 99MBA091A 6 16 c MF 2010C d MF 2017C Unit mm ...
Page 83: ...6 Specifications No 99MBA091A 6 17 e MF 3017C f MF 4020C Unit mm ...
Page 84: ...No 99MBA091A 6 18 6 8 2 Control unit a LED illuminator b Halogen illuminator Unit mm ...
Page 86: ...No 99MBA091A 6 20 d MF 2017C e MF 3017C Unit mm ...
Page 87: ...6 Specifications No 99MBA091A 6 21 f MF 4020C Unit mm ...
Page 88: ...No 99MBA091A 6 22 MEMO ...