No. 99MBA091A
6-14
6.7.6
Connecting QM-Data 200
It is necessary to perform the QM-Data 200 setting when connecting it to the counter unit.
For information about setting up QM-Dada 200, refer to its user’s manual.
(1)
Connection
(2)
QM-Data 200 setting
①
Select the “SYSTEM” key of operation panel.
②
5: Measuring Instrument Setting
-->
1: Counter Type
-->
MF-A/MF-UA
After the selection, the external counter setting of QM-Data 200 becomes the following
communication condition:
●
Baud Rate: 9600 bps
●
Communication Condition: EVEN, 7, 1
●
Flow Control: DTR/DSR
●
Terminator: CR+LF
③
2: Axis Configuration
-->
1
:
X, Y or 2
:
X, Y, Z, (depending on the axis configuration of the
microscope
)
NOTE
When QM-Data 200 starts, the counter display of microscope goes off and LED of
mm/E display at the counter blinks on and off. The switches on the microscope
counter become ineffective.
TIP
A value in units of “E” is displayed by dividing a displacement in units of “mm” by
25.4.
Connecting cable
(
No.12AAA807
)
QM-Data200
Counter
unit
SERIAL(1)
RS232C(2)