1. Overview
No. 99MBA091A
1-13
(3)
Parameter setting operation examples
Parameter
setup example
Button
operation
Counter display
Operating procedure
Setting the
display unit
(Select “mm”)
X
Y
1.Press the X-axis zero-set
button to select parameter
[F01].
2.Press the Y-axis zero-set
button to select [mm].
Setting the count
direction
(Reverse the
count direction
of X-axis)
X
Y
1.Press the X-axis zero-set
button to select parameter
[F02-1].
2.Press the Y-axis zero-set
button to select the setting
(here [dn]) unlike the present
setting (here [UP]).
Setting the count
direction
(Reverse the
count direction
of Y-axis)
X
Y
1.Press the X-axis zero-set
button to select parameter
[F02-2].
2.Press the Y-axis zero-set
button to select the setting
(here [UP]) unlike the present
setting (here [dn]).
Setting the count
direction
(Reverse the
count direction
of Z-axis)
X
Y
1.Press the X-axis zero-set
button to select parameter
[F02-3].
2.Press the Y-axis zero-set
button to select the setting
(here [UP]) unlike the present
setting (here [dn]).
Setting a
Resolution
(Select “0.1
µ
m”)
X
Y
1.Press the X-axis zero-set
button to select parameter
[F03].
2.Press the Y-axis zero-set
button to select [0.1].
Summary of Contents for MF 1010C Series
Page 2: ......
Page 54: ...No 99MBA091A 3 14 MEMO ...
Page 66: ...No 99MBA091A 5 6 MEMO ...
Page 82: ...No 99MBA091A 6 16 c MF 2010C d MF 2017C Unit mm ...
Page 83: ...6 Specifications No 99MBA091A 6 17 e MF 3017C f MF 4020C Unit mm ...
Page 84: ...No 99MBA091A 6 18 6 8 2 Control unit a LED illuminator b Halogen illuminator Unit mm ...
Page 86: ...No 99MBA091A 6 20 d MF 2017C e MF 3017C Unit mm ...
Page 87: ...6 Specifications No 99MBA091A 6 21 f MF 4020C Unit mm ...
Page 88: ...No 99MBA091A 6 22 MEMO ...