No. 99MBA091A
1-12
(2)
Parameter setting procedure
Parameter starting and exiting procedure
Parameter
setup example
Button
operation
Counter display
Operating procedure
①
Starting the
parameter setup
mode
X + Y
+
Power
switch
Turn on the power switch of
microscope main unit while
pressing the X-axis and Y-axis
zero-set buttons at the same
time.
②
Switching
parameters
X
F01→F02-1→・・・・
F11→F01→・・・・
Each time the X-axis zero-set
button is pressed, the parameter
display changes to the next in
incremental order.
The parameters are looped.
③
Switching
parameter
settings
Y
0.1→0.5→1.0→0.1・・・
(
For a resolution
)
Each time the Y-axis zero-set
button is pressed, the parameter
setting changes to the next. The
parameter settings are looped.
④
Exiting the
parameter setup
mode
(
Loading
the set content
)
X + Y
Keep pressing the X-axis and
Y-axis zero-set buttons at the
same time for more than 3
seconds.
Parameter display
(
Y-axis display unit
)
Parameter display
(
X-axis display unit
)
Y
zero-set button
X
zero-set button
Summary of Contents for MF 1010C Series
Page 2: ......
Page 54: ...No 99MBA091A 3 14 MEMO ...
Page 66: ...No 99MBA091A 5 6 MEMO ...
Page 82: ...No 99MBA091A 6 16 c MF 2010C d MF 2017C Unit mm ...
Page 83: ...6 Specifications No 99MBA091A 6 17 e MF 3017C f MF 4020C Unit mm ...
Page 84: ...No 99MBA091A 6 18 6 8 2 Control unit a LED illuminator b Halogen illuminator Unit mm ...
Page 86: ...No 99MBA091A 6 20 d MF 2017C e MF 3017C Unit mm ...
Page 87: ...6 Specifications No 99MBA091A 6 21 f MF 4020C Unit mm ...
Page 88: ...No 99MBA091A 6 22 MEMO ...