5 External Input/Output Function
37
No. 99MBC109B
5.3.2 Communication Commands
This section explains the command format from a PC or PLC, output from the
Counter, and operation details.
Command format
Corresponding output
Operation details
Notes
GA**CRLF
G#**, +01234.567CRLF Outputs "Display value"
See *1
CN**CRLF
CH**CRLF
Switches the display to
"Current value"
See *2
CX**CRLF
CH**CRLF
Switches the display to
"Maximum value"
See *2
CM**CRLF
CH**CRLF
Switches the display to
"Minimum value"
See *2
CW**CRLF
CH**CRLF
Switches the display to
"TIR"
See *2
CR**CRLF
CH**CRLF
Zero setting
CL**CRLF
CH**CRLF
Clears the peak value
CP**,+01234567CRLF CH**CRLF
Inputs the Preset value
See *3
CD**,+01234567CRLF CH**CRLF
Inputs tolerance value S1
See *3
CE**,+01234567CRLF CH**CRLF
Inputs tolerance value S2
See *3
CF**,+01234567CRLF CH**CRLF
Inputs tolerance value S3
See *3
CG**,+01234567CRLF CH**CRLF
Inputs tolerance value S4
See *3
CS**CRLF
CH**CRLF
Clear error
CK**CRLF
CH**,%CRLF
Checks the HOLD status
See *4, *5
*1 "**" denotes an RS-232C Linear Gage channel number (01 to 99) ("00" means all channels). Channels
01 to 04 correspond to CEL1 to CEL4, respectively.
A "#" after "G" in the output data denotes the type of data (N: Current value, X: Maximum value, M:
Minimum value, W: TIR).
*2 If Peak mode is switched using an RS command, data will not be backed up in internal memory.
*3 For the Preset value and tolerance value, enter a value consisting of a +/- sign and an 8-digit of numeric
value without a decimal point.
Perform the tolerance setting in the order CD command → CG command for 3-step tolerance, and in
the order CD command → CE command → CF command → CG command for 5-step tolerance.
An error will be output in the following cases. In this case, redo the settings from the CD command.
・
If the order of the tolerance values is incorrect
・
If the step count and sent data are different
・
If incorrect data is sent
*4 The response output from the CK command ("%") shows the HOLD status.
%=0: Normal status/1: HOLD status
*5 If parameter number 28 is 1 (channel synchronization), all Counters that are connected via RS LINK
when the CK command is sent will switch to the HOLD state.
The HOLD state will be canceled by reading out data with the GA command.
The CK command is valid only with channel number 1 (CK01CRLF).