4 Basic Operations
32
No. 99MBC109B
4.7
Optional Constant Value Setting
You can specify an internal counter (CEL1 to CEL4) to which a multiplication
factor is applied and set the factor for the internal counter value.
If this function is used, the accuracy cannot be guaranteed.
Tips
For EH-101P, there is only 1 internal counter (CEL).
4.7.1 Procedure for Specifying an Internal Counter
This procedure explains how to specify the internal counter (CEL1 to CEL4) to
which a multiplication factor is applied.
1
Press and hold [P.SET], and then press [A_ZERO]/[ZERO].
» The Counter enters Parameter mode.
2
Press [A_ZERO]/[ZERO] once to set parameter number 00 to 1.
» The display appears as to the right.
3
Press [P.SET].
» The parameter number will switch to 40.
Tips
For 1-axis models, the parameter number will switch to 41. Proceed to step
5
.
4
Repeatedly press [P.SET] to advance the parameter number to 41.
» The set value of parameter number
41 for CEL1 will be displayed.
Parameter number
CEL
number
Set value
Tips
To specify another CEL, repeatedly press [P.SET]. The CEL number will advance.
5
Repeatedly press [A_ZERO]/[ZERO] to set the set value to 4.
» The value will be set to 4. (The Constant value will be set to "Option".)
6
Press and hold [P.SET], and then press [A_ZERO]/[ZERO].
» The Counter will return to the stand-by state.