29
GB
6DWLVI\WKHPLQLPXPLQVHUWLRQGHSWKLQWKHWDEOHEHORZZKHQH[SDQGLQJ¿HOG
supplied piping
Pipe diameter
PP>LQ@
Minimum insertion depth
PP>LQ@
>@RUPRUHOHVVWKDQ>@
>@
>@RUPRUHOHVVWKDQ>@
>@
>@RUPRUHOHVVWKDQ>@
>@
>@RUPRUHOHVVWKDQ>@
>@
>@RUPRUHOHVVWKDQ>@
>@
>@RUPRUHOHVVWKDQ>@
>@
$IWHUHYDFXDWLRQDQGUHIULJHUDQWFKDUJLQJHQVXUHWKDWWKHKDQGOHLVIXOO\
RSHQ,INHHSWKHYDOYHFORVHGWKHKLJKRUORZSUHVVXUHVLGHRIWKH
UHIULJHUDQWFLUFXLWPD\EHVXEMHFWHGWRDEQRUPDOSUHVVXUHDQGPD\GDPDJH
WKHFRPSUHVVRUIRXUZD\YDOYHHWF
'HWHUPLQHWKHDPRXQWRIDGGLWLRQDOUHIULJHUDQWFKDUJHE\XVLQJWKHIRUPXOD
DQGFKDUJHUHIULJHUDQWWKURXJKWKHVHUYLFHSRUWDIWHUFRQQHFWLQJWKHSLSHV
7LJKWHQWKHVHUYLFHSRUWDQGFDSVHFXUHO\VRJDVGRHVQRWOHDN5HIHUWRWKH
WDEOHEHORZIRUDSSURSULDWHWLJKWHQLQJWRUTXH
Appropriate tightening torque:
Outside
GLDPHWHURI
copper pipe
PP>LQ@
Cap
1ÂPNJÂFP
6KDIW
1ÂPNJÂFP
6L]HRI
KH[DJRQDO
wrench
PP
Service port
1ÂPNJÂFP
¡>@
¡>@
¡>@
¡>@
¡>@
Caution:
.HHSWKHYDOYHFORVHGXQWLOWKHUHIULJHUDQWFKDUJLQJLV¿QLVKHG
Opening the valve before charging the refrigerant may cause damage
to the unit.
Do not use a leak detection additive.
10.3. Airtight test, evacuation, and
refrigerant charging
1
Airtight test
3HUIRUPZLWKWKHYDOYHRIWKHKHDWVRXUFHXQLWFORVHGFKDUJLQJWKH
UHIULJHUDQWSUHVVXUL]HWKHFRQQHFWLRQSLSLQJDQGWKHLQGRRUXQLWIURPWKH
VHUYLFHSRUWSURYLGHGRQWKHYDOYHRIWKHKHDWVRXUFHXQLW$OZD\VSUHVVXUL]H
IURPERWKWKHKLJKSUHVVXUHJDVSLSHDQGWKHORZSUHVVXUHOLTXLGSLSH
VHUYLFHSRUWV
[Fig. 10.3.1] (P.12)
A
Nitrogen gas
B
To indoor unit
C
System analyzer
D
Low knob
E
High knob
F
Valve
G
/RZSUHVVXUHOLTXLGSLSH
H
+LJKSUHVVXUHJDVSLSH
I
Heat source unit
J
Service port
2EVHUYHWKHIROORZLQJUHVWULFWLRQVZKHQFRQGXFWLQJWKHDLUWLJKWQHVVWHVW
WRSUHYHQWQHJDWLYHHIIHFWVRQWKHUHIULJHUDWLQJPDFKLQHRLO
Also, with
QRQD]HRWURSLFUHIULJHUDQW5$JDVOHDNDJHFDXVHVWKHUHIULJHUDQW
FRPSRVLWLRQWRFKDQJHDQGDIIHFWVSHUIRUPDQFH7KHUHIRUHFRQGXFWWKH
airtightness test cautiously.
Caution:
Only use R410A refrigerant.
8VLQJRWKHUUHIULJHUDQWVVXFKDV5RU5&ZKLFKFRQWDLQVFKORULQH
ZLOOGHWHULRUDWHWKHUHIULJHUDWLQJPDFKLQHRLORUFDXVHWKHFRPSUHVVRUWR
PDOIXQFWLRQ
2
Evacuation
(YDFXDWHZLWKWKHYDOYHRIWKHKHDWVRXUFHXQLWFORVHGDQGHYDFXDWHERWKWKH
FRQQHFWLRQSLSLQJDQGWKHLQGRRUXQLWIURPWKHVHUYLFHSRUWSURYLGHGRQWKH
YDOYHRIWKHKHDWVRXUFHXQLWXVLQJDYDFXXPSXPS$OZD\VHYDFXDWHIURP
WKHVHUYLFHSRUWRIERWKWKHKLJKSUHVVXUHJDVSLSHDQGWKHORZSUHVVXUH
OLTXLGSLSH$IWHUWKHYDFXXPUHDFKHV3D>DEV@>SVL7RUU@
,
HYDFXDWHIRUDWOHDVWRQHKRXURUPRUH7KHQVWRSWKHYDFXXPSXPSDQG
OHDYHLWIRUKRXU(QVXUHWKHGHJUHHRIYDFXXPKDVQRWLQFUHDVHG
(If the
degree of vacuum increase is larger than 130 Pa [0.01886 psi/1.0 Torr],
water might present. Apply pressure to dry nitrogen gas up to 0.05
MPa [7.25 psi] and vacuum again. Repeat the evacuation process three
or more times until the vacuum pressure is lost by 130 Pa or below.)
)LQDOO\VHDOLQZLWKWKHOLTXLGUHIULJHUDQWWKURXJKWKHKLJKSUHVVXUHJDVSLSH
DQGDGMXVWWKHORZSUHVVXUHOLTXLGSLSLQJWRREWDLQDQDSSURSULDWHDPRXQWRI
WKHUHIULJHUDQWIRURSHUDWLRQ
1HYHUSHUIRUPDLUSXUJLQJZLWKUHIULJHUDQW
[Fig. 10.3.2] (P.12)
A
System analyzer
B
Low knob
C
High knob
D
Valve (heat source
XQLW
E
/RZSUHVVXUHOLTXLG
pipe
F
+LJKSUHVVXUHJDV
pipe
G
Service port
H
7KUHHZD\MRLQW
I
Valve
J
Valve
K
5$F\OLQGHU
L
Scale
M
Vacuum pump
N
To indoor unit
O
Heat source unit
Note:
Always add the appropriate amount of refrigerant. Also always charge
the system with liquid refrigerant.
Use a gauge manifold, charging hose, and other parts for the
refrigerant indicated on the unit.
Use a graviometer. (One that can measure down to 0.1 kg [302 oz].)
8VHDYDFXXPSXPSZLWKDUHYHUVHÀRZFKHFNYDOYH
(Recommended vacuum gauge: ROBINAIR 14830A Thermistor Vacuum
Gauge or Micron Gauge)
Do not use a manifold gauge to measure vacuum pressure.
Also use a vacuum gauge that reaches 65 Pa [abs] [0.00943 psi/0.5
7RUU@RUEHORZDIWHURSHUDWLQJIRU¿YHPLQXWHV
<Triple Evacuation>
Evacuate the system to 4,000 microns from both service valves.
System manifold gauges must not be used to measure vacuum.
A micron gauge must be used at all times.
%UHDNWKHYDFXXPZLWK1LWURJHQ1LQWRWKHGLVFKDUJHVHUYLFHYDOYHWR
PSIG.
Evacuate the system to 1,500 microns from the suction service valve.
%UHDNWKHYDFXXPZLWK1LWURJHQ1LQWRWKHGLVFKDUJHVHUYLFHYDOYHWR
PSIG.
Evacuate the system to 500 microns. System must hold the vacuum at
500 microns for a minimum of 1 hour.
Conduct a rise test for a minimum of 30 minutes.
3
Refrigerant Charging
Do not use refrigerant other than the type indicated in the manuals
provided with the unit and on the nameplate.
'RLQJVRPD\FDXVHWKHXQLWRUSLSHVWREXUVWRUUHVXOWLQH[SORVLRQRU¿UH
GXULQJXVHGXULQJUHSDLURUDWWKHWLPHRIGLVSRVDORIWKHXQLW
,WPD\DOVREHLQYLRODWLRQRIDSSOLFDEOHODZV
0,768%,6+,(/(&75,&&25325$7,21FDQQRWEHKHOGUHVSRQVLEOH
IRUPDOIXQFWLRQVRUDFFLGHQWVUHVXOWLQJIURPWKHXVHRIWKHZURQJW\SHRI
UHIULJHUDQW
Airtight test procedure
Restriction
$IWHUREWDLQLQJWRWKHGHVLJQSUHVVXUH03D>SVL@XVLQJQLWURJHQJDVDOORZ
WKHV\VWHPWRVWDQGIRUDERXWRQHGD\,IWKHSUHVVXUHGRHVQRWGURSDLUWLJKWQHVVLV
good.
+RZHYHULIWKHSUHVVXUHGURSVDQGWKHOHDNDJHDUHDLVXQNQRZQWKHIROORZLQJEXEEOH
WHVWPD\DOVREHSHUIRUPHG
$IWHUWKHSUHVVXUL]DWLRQGHVFULEHGDERYHVSUD\WKHÀDUHFRQQHFWLRQSDUWVEUD]HG
SDUWVDQGRWKHUSRWHQWLDOOHDNDJHDUHDVZLWKDEXEEOLQJDJHQW.\XERÀH[HWFDQG
YLVXDOO\FKHFNIRUEXEEOHV
$IWHUWKHDLUWLJKWWHVWZLSHRIIWKHEXEEOLQJDJHQW
,IDÀDPPDEOHJDVRUDLUR[\JHQLVXVHGDVWKHSUHVVXUL]DWLRQ
JDVLWPD\FDWFK¿UHRUH[SORGH