The Main Unit
1.Belt
2.LCD Display
3.LOGO
4. Hooks
5. Menu Keys
6. Alarm LED
7. Power LED
8. Keypad
9. Rotary Knob
10. Product Label
11.Support Pillar Screw
12. Battery Case
13. Power adapter port
14. Battery Switch
15. USB Socket
16., Probe Cable Port (Transmit)
17. Probe Cable Port (Receive)
2.2 Screen Display
The instrument displays are designed to be easy to interpret.
The Instrument has a
digital screen for the
display of A-Scan.
The status line below
the wave display shows
values of range settings
and measured values.
On the first status line
Echo height is displayed as: 24.5%. It means that the peak echo height inside the current selected
gate is 24.5% FSH (Full Screen Height of the wave area).
Flaw size value means ERS (Equivalent Reflector Size) of the reflector signal inside the current
selected gate. It is displayed after the AVG curve is switched on. When the DAC curve is on, this
value will be displayed as the dB offset of the peak echo inside the current selected gate to the
reference DAC curve (DAC
→
OFFSET
→
SIZE REF).
The icon
‘ ’
indicates that the horizontal scale is displayed as sound path (S-PATH). It can be
changed to
‘ ’
(Projection value) or
‘ ’
(Depth) by simply pressing
repeatedly.