![MITECH MFD500B User Manual Download Page 11](http://html.mh-extra.com/html/mitech/mfd500b/mfd500b_user-manual_1806972011.webp)
10
Connect the instrument to the mains socket-outlet using the power supply unit. The plug receptacle is at
the top left of the instrument. Push the plug of the power supply unit into the plug receptacle until it snaps
into place with a clearly audible click. The PWR LED on the keypad of the instrument will light in green
color if the connection is properly aligned
Operation Using Batteries
Use a lithium-ion battery pack provided with the instrument for the battery operation.
The battery compartment is situated at the instrument back. The cover is fastened with 4 screws To
insert the battery pack
Loosen the four screws of the battery cover.
Lift the cover off upward.
Insert the battery into the battery compartment.
Close the battery compartment and fasten the screws.
Turn on the instrument to make sure the battery is installed correctly and firmly.
Note:
The battery switch on the top of the instrument must be set to ON when operating with the battery
pack or charging it.
When not using the instrument, set the switch to OFF to save power.
3.2 Connecting a Probe
To prepare the instrument for operation, you have to connect a probe to it. The instrument is available
with the probe connectors BNC (LEMO connectors are optional).
When connecting a probe to the instrument, it’s not only important that the probe’s physical connection
be properly made. It’s also important that the instrument is properly configured to work with the installed
probe. The Instrument will operate with one or two single-element probes or with a dual-element probe.
To install a single-element probe, connect the probe cable to either of the two ports on the front of the
instrument. When two probes or a dual-element probe is connected to the instrument, the “Receive”
probe connector should be installed in the right port and the ‘Transmit’ probe connector in the left port.
3.3 Starting the Instrument
To start the instrument, press the switch-on key
. If it operates on the internal battery pack, make sure
to set the battery switch to ON position before starting.
The start display of the instrument appears; here you will also see the current software version and the
serial number of the instrument. The instrument carries out a self-check and then switches over to
stand-by mode.
The settings of all function values are the same as before switching-off of the instrument.
The instrument will shut off automatically when the battery capacity level is too low.
4 Operation
4.1 Adjusting the Display Range
The function group BASE enables you to make the basic adjustment of the display range. The display on
the screen must be adjusted for the material to be tested and for the probe used.
Calibration requires the use of calibrated standard made of the same material as the test piece. Prior to
calibrating the instrument/probe combination, the A-Scan display-screen range (the material thickness
Connect one single element
probe to either port.
Connect leads from a Dual
Element Probe to both ports.
For through-transmission,
connect two single element
probes to the transmit (left,
labeled as
) and receive
(right, labeled as
) ports.