94
2. OUTPUT DATA (PC
→
Instrument)
1. Measurement Data
Measurement data is output by the CA-100Plus according to the present display mode. However, the format of the
output data will be slightly different from the format of the displayed data.
When the optional 4-point expansion board CA-B04 is used
The probe connector number for which data will be output should be selected using the command:
“ O probe connector number ”
In the following, “_” indicates a space.
●
Display mode: xyLv
Format: probe connector number _ x value ; y value ; Lv value delimiter
Probe connector number: Value from 1 to 5; probe connector number for which data is being output. In example,
data is being output for probe connector P1.
x value, y value: First three decimal places output. In example, x = 0.310 and y = 0.330.
Lv value: Up to four characters output; if value consists of fewer than four characters, value will be preceded by the
required number of spaces. In example, Lv = 150.
•
Example: P1 _ 310; 330; _ 150 delimiter
●
Display mode: T
∆
uvLv
Format: P probe connector number _ T value ;
∆
uv value ; Lv value delimiter
Probe connector number: Value from 1 to 5; probe connector number for which data is being output. In example,
data is being output for probe connector P1.
T value: Up to five characters output; if value consists of fewer than five characters, value will be preceded by the
required number of spaces. In example, T = 6500.
∆
uv value: First three decimal places output; preceded by a space if positive or “-” if negative. In example,
∆
uv =
+ 0.005.
Lv value: Up to four characters output; if value consists of fewer than four characters, value will be preceded by the
required number of spaces. In example, Lv = 150.
•
Example: P1 _ _ 6500; _ 005; _ 150 delimiter
●
Display mode: Analyzer (RBG)
Analyzer (RBG) display mode is available only with an optional Analyzer Card.
Format: P probe connector number _ R value ; B value ; G value delimiter
Probe connector number: Value from 1 to 5; probe connector number for which data is being output. In example,
data is being output for probe connector P1.
R value, B value, G value: Up to five characters output; if value consists of fewer than five characters, value will be
preceded by the required number of spaces. In example, R = 101, B = 99.0, and G = 114.
•
Example: P1 _ _ _ 101; _ 99.0 _ _ 114 delimiter
When the optional 4-point expansion board CA-B04 is used
Data output for more than one probe connector number can be selected if desired. If more than one probe connector
number is selected, data will be output consecutively for all selected probe connector numbers in numerical order.
Format: P probe connector number _ measurement data* delimiter
P probe connector number _ measurement data* delimiter
...
.
P probe connector number _ measurement data* delimiter
∗
Measurement data is output ac-
cording to the format above for
the selected display mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for CA-100PLUS -
Page 25: ...22 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 47: ...44 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 101: ...98 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 124: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...