39
Measur
ement Pr
epar
ation
SELECT : SYNC.
4. INT
6
0.0Hz
<Changing the Synchronizing Frequency for INT>
Select INT as explained above, and enter the desired frequency using the ten-key.
<Relationship between Measurement Speed and SYNC Mode>
The measurement time (sampling time) is determined by the selected SYNC mode.
The measurement speed (the number of measurements and outputs per second) is determined by the SYNC mode
and the following conditions.
•
Intensity of the display to be measured
•
Measurement mode
•
Data output (RS-232C or USB)
•
Baud rate when data is output via RS-232C
•
Number of measuring probes to be used (when the optional 4-point expansion board is used)
The table below shows the measurement speed for each SYNC mode when measurement is performed under the
following conditions.
RS232C
•
Intensity of the display to be measured ... No errors and warnings, and the luminance is stable.
•
Display mode ........................................... xyLv
•
Baud rate for data output ......................... 38,400 bps
•
Number of connected measuring probes .... 1
xyLv
FAST
SLOW
NTSC
17 measurements/sec.
4.5 measurements/sec.
PAL
15 measurements/sec.
4 measurements/sec.
EXT*
17 measurements/sec.
4.5 measurements/sec.
UNIV.
7 measurements/sec.
1.5 measurements/sec.
INT*
17 measurements/sec.
4.5 measurements/sec.
* The measurement speed for EXT and INT when the vertical scanning frequency is 60 Hz is given.
USB
•
Intensity of the display to be measured ... No errors and warnings, and the luminance is stable.
•
Display mode ........................................... xyLv or flicker
•
Number of connected measuring probes .... 1
xyLv
FAST
SLOW
NTSC
20 measurements/sec.
5 measurements/sec.
PAL
17 measurements/sec.
4 measurements/sec.
EXT*
20 measurements/sec.
5 measurements/sec.
UNIV.
8 measurements/sec.
1.5 measurements/sec.
INT*
20 measurements/sec.
5 measurements/sec.
* The measurement speed for EXT and INT when the vertical scanning frequency is 60 Hz is given.
4
.
Press the
key to confirm the selection.
* To use EXT mode, the vertical synchronizing signal used for the display must be input to the instrument. (Page 28)
* By default (factory setting), the instrument is set so that EXT mode will be selected automatically when the POWER switch is set to ON.
If you want to change this setting, refer to page 29.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for CA-100PLUS -
Page 25: ...22 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 47: ...44 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 101: ...98 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 124: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...