81
Measur
ement Section
[Operating Procedure]
1
.
Set the POWER switch to ON.
2
.
Place receptor area of measuring probe
face down on a flat surface so that no light
reaches the receptor area.
Never direct the measuring probe toward a high-inten-
sity illuminant.
When the optional 4-point expansion board CA-B04 is used
Block the receptor of each measuring probe from light.
If there are any receptors not blocked from light, zero
calibration will not be performed correctly.
3
.
Press the 0-CAL key.
4
.
Press the MODE key to select analyzer
measurement mode (RGB).
5
.
Press the MEMORY CH
and
keys
to select the memory channel for which the
RGB emission characteristic for analyzer
mode has been set (page 59).
When the optional 4-point expansion board CA-B04 is used
Select the probe no. for which the RGB emission characteristic for the analyzer mode has been input.
1
1
1
1
1
Press the
key.
The LCD display section will switch to the menu selection screen.
2
2
2
2
2
Press the
key to open the PROBE selection screen.
Each time the
key is pressed, the screen will switch in the
order PROBE
→
SYNC
→
ID Name input
→
RANGE
→
Mea-
surement Speed
→
Number of Digits
→
Calibration Standard
→
RS232C Baud Rate
→
PROBE.
3
3
3
3
3
Press the
key to display the probe no. you want to select.
Each time the
key is pressed, the probe no. switches in the
order [P1] ….
4
4
4
4
4
Press the
key to confirm the selection.
* By default (factory setting), the instrument is set so that [P1] will be selected automati-
cally when the POWER switch is set to ON. If you want to change this setting, refer to
page 29.
6
.
Place the measuring probe against the display and
take measurement.
1
3
2,3
6
4
5
5
7
DARKEN PROBE
PUSH 0-CAL KEY
ZERO CALIBRATION
CH00 EXT Ad P1
[ ]
CH01 EXT P3
[ ]
MENU : SELECT
PUSH SPACE KEY
SELECT : PROBE
P1 35881112
SELECT : PROBE
P3 35881113
Place the probe against the display
Message displayed
when the POWER
switch is set to ON
Block entry of light
Press the 0-CAL key.
During zero calibration
End of zero calibration
Memory channel
Probe no.
Menu selection screen
PROBE selection screen
Probe no.
Press the
key until the
desired probe
no. appears.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for CA-100PLUS -
Page 25: ...22 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 47: ...44 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 101: ...98 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 124: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...