According to the procedures in the section 4, check on
various kinds of connecting conditions. Especially, check
that if there may have effects from the noise source such
as inverter or like that.
4
Same condition
YES
NO
Voltage and current
output is stable
NO
YES
YES
NO
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details.
Remove the connecting cable for strain gage applied transducer from the terminals
Set the connecting range to DC
・
V for the measuring instrument such as tester
and so on.
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details.
Short between B
−
D and V
−
on
the terminals
Start of
measurement
The voltage between
A and C is just like the setting
of DIP2(refer to 6
−
3), and
stable as well