Calculation example
)
This is the calculation example for using 1 point
load cell with rated capacity of 1 t and rated
output of 3 mV/V, make the voltage output of
10.000 V for the maximum load of 0.34 t
Warning
:
The rated output value applied
on calculation should be the same value as
described on the inspection data of each load
cell.
Rated load value of load cell
1 t
Rated output value of load cell
3 mV/V
The maximum load value
0.34 t
Voltage output value at the maximum load
10.000 V
CALIB value
1 mV/V
(Load cell output value at the maximum load)
(Rated output value of load cell)
×
(The maximum load value)
(Rated load value of load cell)
3 mV/V
×
0.34 t
1 t
1.02 mV/V
(Voltage output value equivalent to CALIB value)
(Voltage output value at the maximum load)
×
(CALIB value)
(Load cell output value at the maximum load)
10.000 V
×
1 mV/V
1.02 mV/V
9.804 V
Consequently, the voltage output value in
turning ON the “CAL” switch should be adjusted
to 9.804 V by using “SPAN” trimmer.
=
=
=
=
≒
=
4
Zero adjustment
Turn OFF the CAL switch, then confirm the
voltage output value at 0.000 V, or the current
output value(option) at 4.000 mA. Otherwise,
return to the step 2.
5
Now, calibration has completed.