3. Connections
4
−
3
−
1. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure
transducer and so on. Here, we will describe the example of connections with load cell, so
connect with another type of strain gage applied transducers in the same way.
※
1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and output of “
+
”
direction is required, connect “Green” with B terminal, and “Blue”
with D terminal individually. As there is a case which standard
wiring color is different, please confirm the inspection data sheet of
the load cell being used.
※
2 When the total length of CAB
−
502 specifies more than 30 m, the
accuracy may be out of warranty because the resistance of cable
makes the input voltage of the instrument decreased.
※
3 When the length of CAB
−
502 cable is applied more than 10 m, or
when the system is using the zener barrier, the CALIB is not
applicable.
1
Connection with one set of load cell and CSA
−
507C
3 m
RED
※
1 GRN
WHT
(Shield)
※
1 BLU
YEL