7
※
1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and output of “
+
”
direction is required, connect “Green” with B terminal, and “Blue”
with D terminal individually. As there is a case which standard
wiring color is different, please confirm the inspection data sheet of
the load cell being used.
※
2 When the total length of CAB
−
502 specifies more than 30 m, the
accuracy may be out of warranty because the resistance of cable
makes the input voltage of the instrument decreased.
※
3 When the length of CAB
−
502 cable is applied more than 10 m, or
when the system is using the zener barrier, the CALIB is not
applicable.
2
Connection with one set of load cell and junction box for extension use B
−
304 and CSA
−
507C
Junction Box
B
−
304
CAB
−
502
※
2
(
Length of CAB
−
502 should be
within 30 m totally.
)
3 m
RED
WHT
(Shield)
YEL
※
1 GRN
※
1 BLU
B
−
304 internal terminal connection
to CSA
−
507C
to load cell
RED
WHT
BLU
GRN
YEL(Shield)
Terminal pitch 9.5 mm
Suitable crimp
−
type
terminal lugs
:
1.25
−
4 or 2
−
4
RED
WHT
BLU
GRN
YEL(Shield)