3. Connection with power supply and ground
Connection with power supply and ground should be made as follows:
Beside, grounding should be the D class with single ground.
Power supply voltage
AC100 V (AC90 V
〜
AC115 V)
Power supply frequency 50/60 Hz
Power consumption
Approx. 5 VA
(
without any options at AC100 V
)
〜
AC100 V
(AC90 V
〜
115 V)
D class single ground
※
When the power supply is AC200 V(CSA507C
−
P63), refer to the paragraph 7
−
3.
Warning
●
Connection with power supply and ground should be made securely as
the figure indicates and also should be used within the specified
condition of power supply.
If neglected, it may cause an unexpected failure.
●
Grounding for the instrument should be the D class with single
ground. If neglected, it may cause an unexpected malfunction due to
the effects of noise from the other equipments.
●
GND terminal is connected with the frame of the instrument.