MP 26/10
Installation and Calibration Manual
Minebea Intec
EN-35
6.3.
Detection & Display of Sensor error & Wiring errors
1)
Break in the supply measuring or sense lines: "
FAIL
" is displayed on the process value
display and “
FbF.1
" in the error list.
(A substitute value (In.F ) can be entered in the event of a sensor error.)
2)
Wrong polarity of supply, sense and input signal lines: "
FAIL
" appears on the process value
display and "
POL.1
" in the error list. Detection of wrong polarity of sense takes place
during instrument start-up.
Subsequent correction of the sense wiring is detected only after instrument re-start (i.e.
the instrument must be switched off and on again.)
3)
Short circuit of supply and sense lines ==> behavior as with break in supply lines: "
FAIL
"
appears on the process value display and "
FbF.1
" in the error list.
4)
Short circuit of measuring lines: measuring signal = 0
5)
Overshoot of measurement input: “
FAIL
” appears on the process value display and
"
FbF.1
" in the error list.