MP 26/10
Installation and Calibration Manual
Minebea Intec
EN-9
2.5.
Connecting output OUT3
The Out3 is an analog output through which voltage or current can be led
based on the configuration.
Current (0...20mA)
ports: 11, 12
Voltage (0...10V)
ports: 12, 13
Its current output is configurable 0/4...20 mA, short-circuit proof.
Control range:
-0.5....23 mA
Load:
<
600 Ω
Load effect:
<
0.02%
Resolution:
<
1.5 µA
Error:
<
0.1%
Its voltage output is configurable 0/2...10V, not permanently short-circuit proof
Control range
0.15...11.5 V
Load:
>
2 kΩ
Load effect:
<
0.06%
Resolution:
<
0.75 mV
Error:
<
0.1%
Additional error when
using the current output
simultaneously
<
+0.09%
2.6.
MP 26/10 software
2.6.1.
Installation
The software tool for MP 26 transmitter is contained in the CD-ROM; it can run under
Windows 98, -NT, – 2000 or – XP. The steps for installing the software are given below.
1.
Switch off the PC and connect the RS-232 cable from transmitter to the PC which is
provided along with the device.
2.
Start the PC.
3.
Ensure that the necessary administrator rights are given to install executable
programs.
4.
Double click the Setup.exe.
5.
Follow the instructions given.
6.
Select the destination directory (e.g. C:\Programs\Minebea-Intec\Mp26).
7.
After successful installation the message: Installation finished will appear.
8.
Start the program MP26/10