MP 26/10
Installation and Calibration Manual
Minebea Intec
EN-22
4.3.2.
Calibration by Weight
1.
Load the data from device.
2.
Confirm the settings.
3.
Go to view parameter as shown in the figure below.
Set the dead load/span:
The dead load is the constant load which is always present on
the weighing scale/system.
4.
Remove any weight apart from dead load from the system. Then enter OuL.1 as zero in
parameter
input 1.
5.
Go to view
operations, which displays a window as shown in RHS.
6.
In the operation window note the measured value under meas.value correction
input.1.
7.
Next enter the noted measure value in parameter
input.1
InL.1.
8.
Thus
InL.1 = Measured Value under dead load
OuL.1 = 0