The battery indicator on the operating panel will be unlit if the device is mains powered or lit
green if using battery power.
Battery type
Rechargeable lithium-ion
Run time, when powered by a new fully
charged battery
≥ 500 Auto measurement reports
or 2 hours continuous recording
or 6 hours of measurement without recording
Charge Time (with power off)
≤ 3 hours to 90% capacity
≤ 3.5 hours to 100% capacity
Shutdown delay (after a low battery message
first occurs)
Minimum 5 minutes
•
The battery is charged whenever the device is connected to an AC power source, regardless of
whether the device is switched on.
•
On-screen battery symbols (top right corner of display) indicate battery status as follows:
•
Exchange or conditioning of the battery should be carried out by authorised personnel only.
Please refer to the
BeneHeart R3 Operators Manual
for more information.
Indicates that the battery works correctly. The solid green portion represents the current
battery charge level. Each block represents a charge of approximately 20% capacity.
Indicates that the battery has a low charge level and needs to be charged. In this case, the LED
turns yellow and the message [
Low Battery
] is shown at the bottom of the screen.
Indicates that the battery is almost depleted and needs to be charged immediately
Indicates that no battery is installed, or the battery fails to charge
Section 2
Battery