Title
Loading Paper
•
Before printing ECG reports, ensure that thermal recording paper is loaded. The equipment supports
both roll paper and Z-fold paper.
•
The setting of [
PaperType
] in the [
Maintenance
] menu must correspond with the paper used,
otherwise the system will prompt “Paper Type Error”. This should be done on initial setup of the
device as a password is required. Speak to your Medical Engineering Department for more
information.
•
The default setting is [
Roll
] and this is generally the preferred option as there is little risk of a paper
jam.
Inside the Printer Compartment
To Load Roll Paper
Paper roll
holder
Paper jam protector
Thermal print head
A.
Press down the recorder door latch to open the recorder door.
B.
Check that the paper jam protector is secured flat against the
recorder door (i.e. not clipped into the paper roll holder).
C.
Insert a new paper roll into the paper roll holder, with the grid
side facing the thermal print head.
D.
Unroll the beginning of the paper and close the recorder door.
Overlap the unrolled paper on the recorder door. Make sure
the grid side is facing up.
Section 4
Device Setup