These guidelines do
NOT
replace the instructions for use and all users should refer to the appropriate
Operator’s Manual
for detailed instructions.
The Illustrations may not necessarily reflect the setup or data displayed on your device.
BeneHeart R3 Clinical Education Pack SWL 02.07_V1 Mindray UK © 2020. Errors and omissions excepted. Subject to change.
Mindray® is a registered trademark of Shenzhen Mindray Bio-Medical Electronics Co. Ltd.
23
Paper jammed or
misaligned
1.
Unapproved paper is used
2.
Recording paper is not properly
loaded
1.
Use approved recording paper.
Take out the paper and tear off the
jammed part.
2.
Reload the paper as described in
Section 4: Device Setup
Partially missing
printout or printout not
clear.
1.
Dirty print head
2.
Some thermal points on print
head are damaged
1.
Clean the print head as described
in
Section 15: Care and
Maintenance
2.
If the problem persists, contact
your service personnel
The equipment
automatically shuts
down
1. Auto shutdown is enabled
2. The battery is depleted when
the equipment runs on battery
power
1.
Check the setting of [
Auto Shut
Down
] by selecting
[
Menu
]
→
[
Basic Setup
]. If any limb
lead is detached, the equipment
will automatically shut down when
there has been no operation within
the preset time.
2.
Connect the equipment to AC
mains to run the equipment and
charge the battery
The display is
completely blank
1.
Auto Standby is enabled
1.
If a limb lead is detached and there
has been no operation for the
defined time, the device will
automatically turn off the display to
save power. To exit the standby
mode, press any key.
The screen display is
too dark to be seen
clearly
1.
The setting of brightness is low 1.
Adjust screen brightness