Setting the Waveform
On the main screen:
•
Repeatedly press [
F1
] to select the current waveform speed.
Default Setting: 25mm/s
•
Repeatedly press [
F2
] to select the current waveform gain.
Default Setting: 10mm/mV
•
Press the [
Filter
] (number 5) key to set the the current frequency of muscle artefact filter.
Default Setting: 35 Hz
•
Press the [
Lead
] (number 8) key to select the leads for manual recording.
Default Setting: I, II, III
You can also configure the ECG waveforms by accessing the [
Wave Setup
] menu, however be aware
that any changes made in the menus will be retained when the device is switched off, whilst changes
made with function keys will be reset to the default configuration.
Setting ECG Report
The format and content of the ECG report is configurable in the [
Report Setup
] menu. Please refer to
the
BeneHeart R3 Operators Manual
for a full list of options.