Some patient information may directly affect ECG analysis. Complete and correct patient information is
helpful for accurate diagnosis and treatment of the patient.
For a new patient, enter patient information before taking an ECG measurement.
•
You can quickly enter patient information by pressing the [
ID
] key, [
Age
] key, or [
Gender
] key.
Typing Numbers
•
Use the numerical keys on the operating panel.
•
Use [
X
] on the operating panel to delete.
Typing Letters
•
Select the relevant field i.e. “Last Name” to activate the on-screen keyboard.
•
Use [
F3
] and [
F4
] to navigate the keyboard.
•
Press [
F4
] Enter to select a letter.
•
Press [
Del
] to delete a letter.
•
Press [
OK
] to move onto the next field.
•
When all required information is complete press either [
F1
] Back or [
ECG
] to start analysis.
The device will prompt the user to input patient details when the [
ECG
] button is pressed.
Please see
Set Patient Info
for setup details.
The paediatric lead placement V4R is recommended if the patient is under 16 years of age. Please
record V4R using the V3 electrode (see
Patient Preparation
). This is normal practice for a patient of
this age. This requires the [
V3 Placement
] setting to be changed to [
V4R Position
] in the
[Patient
Info Setup
] menu.
The device will discharge a patient when switched off. However, if you are moving directly to another
patient please ensure the device is cleaned and the new patient’s details are entered by selecting [
ID
],
[
Age
] or [
Gender
].
Section 8
Patient Demographics