BeneHeart R3 

Clinical Education Pack

 

 

 

 

 

Training Excellence

 

Summary of Contents for BeneHeart R3

Page 1: ...www mindrayuk com BeneHeart R3 Clinical Education Pack Training Excellence...

Page 2: ...and omissions excepted Subject to change Mindray is a registered trademark of Shenzhen Mindray Bio Medical Electronics Co Ltd 2 Section Title Page Section Safety Notes 3 1 Battery 4 2 Equipment Overvi...

Page 3: ...ntended to be used with high frequency surgical units Keep away from fluid If liquids have entered the equipment remove the device from use and have it checked by service personnel For paced patients...

Page 4: ...capacity 3 5 hours to 100 capacity Shutdown delay after a low battery message first occurs Minimum 5 minutes The battery is charged whenever the device is connected to an AC power source regardless of...

Page 5: ..._V1 Mindray UK 2020 Errors and omissions excepted Subject to change Mindray is a registered trademark of Shenzhen Mindray Bio Medical Electronics Co Ltd 5 Front View Side Views Back View Bottom View D...

Page 6: ...d otherwise the system will prompt Paper Type Error This should be done on initial setup of the device as a password is required Speak to your Medical Engineering Department for more information The d...

Page 7: ...e diagram can be seen on the trunk cable Plug the patient cable to the connector on the right side of the equipment Ensure the connector on the cable is arrow side up Tighten the screws to securely at...

Page 8: ...F3 to select the desired menu item Press F4 to confirm the selection You can return to previous menu by pressing F1 In the main menu you can Configure waveforms Configure recordings Manage files Cust...

Page 9: ...ect Patient Info Setup Press F4 Enter First select Detailed Patient Info This allows you to decide which details can be included when the Patient ID key is pressed Basic information of ID name gender...

Page 10: ...trademark of Shenzhen Mindray Bio Medical Electronics Co Ltd 10 Waveform Setup No recommended changes to default setup Report Setup Select Extend Record The device automatically performs a rhythm mea...

Page 11: ...password System Setup Functions available Load configuration Export configuration Print configuration Return to default configuration Loading Configuration You can import a configuration file stored o...

Page 12: ...the operating panel to delete Typing Letters Select the relevant field i e Last Name to activate the on screen keyboard Use F3 and F4 to navigate the keyboard Press F4 Enter to select a letter Press...

Page 13: ...air from each electrode site 3 Degrease each electrode site with alcohol and abrade slightly with dry gauze to remove dead skin cells 4 Ensure the skin is completely dry Electrode Positioning Limb ele...

Page 14: ...in a high quality ECG signal make sure that the electrodes firmly contact the skin 3 Route the lead wires to avoid twisting Connect the lead wires with the electrodes 4 Make sure the patient cable is...

Page 15: ...1 Patient ID display 2 Gender icon the display will show or 3 Age the unit can be set to Years Months or Days 4 Heart rate 5 Network status icon 6 USB device status icon 7 System time 8 Battery statu...

Page 16: ...ult Setting 25mm s Repeatedly press F2 to select the current waveform gain Default Setting 10mm mV Press the Filter number 5 key to set the the current frequency of muscle artefact filter Default Sett...

Page 17: ...urement If the Preview option is disabled the equipment automatically prints the ECG report after ECG data is acquired and analysed If the Preview option is enabled the preview of the ECG report displ...

Page 18: ...eforms of the rhythm lead In the rhythm ECG report ECG waveforms are displayed in 3 cascade lines with each line including 10 seconds of waveform You can change the lead to be recorded in the rhythm r...

Page 19: ...lectronics Co Ltd 19 1 Auto Measurement ECG Report 1 Patient information 2 Time of acquisition 3 Global measurements 4 Critical value 5 Diagnosis statement 6 Conclusion Resting 12 Lead ECG Analysis Th...

Page 20: ...ctronics Co Ltd 20 Critical Values Consider Acute STEMI Acute MI Ischemia Extreme Tachycardia Extreme Bradycardia Significant Arrhythmia Prolonged QTc Interval Diagnoses Median Complex Measurement Mat...

Page 21: ...port was originally generated Access the Directory It is possible to access previous reports via the Directory The Directory list will be shown in date and time order with the most recent at the top T...

Page 22: ...ECG data displays unacceptable noise 1 Patient movement 2 Improper filter setting 3 Poor skin preparation 4 Electrode problem 5 Patient cable problem 6 Wrong accessories are used or mix electrode type...

Page 23: ...Some thermal points on print head are damaged 1 Clean the print head as described in Section 15 Care and Maintenance 2 If the problem persists contact your service personnel The equipment automatical...

Page 24: ...s Recommended cleaning agents for the equipment are Sodium hypochlorite bleach diluted Hydrogen peroxide 3 Ethanol 75 Isopropanol 70 For the recommended cleaning agents for the reusable accessories re...

Page 25: ...n swabs 6 When the print head is completely air dried reload the recording paper and close the recorder door Printing quality may be affected by a dirty print head Clean the print head once per month...