A - 4
CAUTION
•
If the equipment is connected through the wireless network in low strength signal, the battery
standby life will be shortened.
A.8
Data Storage
Remaining charge after “Low
Battery” is prompted
For BeneHeart C1/BeneHeart C1A/BeneHeart C1 Fully Automatic/
BeneHeart C1A Fully Automatic/BeneHeart S1/BeneHeart S1A/BeneHeart
S1 Fully Automatic/BeneHeart S1A Fully Automatic:
•
At least 30 minutes operating time (at 20 °C± 5 °C of ambient
temperature, wireless function off, voice volume set to low) and at
least 10 200J discharges (with one minute of CPR between discharges)
•
At least 30 minutes operating time (at 20 °C± 5 °C of ambient
temperature, wireless function off, voice volume set to low) and at
least 6 360J discharges (with one minute of CPR between discharges)
For BeneHeart C2/BeneHeart C2A/BeneHeart C2 Fully Automatic/
BeneHeart C2A Fully Automatic/BeneHeart S2/BeneHeart S2A/BeneHeart
S2 Fully Automatic/BeneHeart S2A Fully Automatic:
•
At least 30 minutes operating time (at 20 °C± 5 °C of ambient
temperature, wireless function off, voice volume set to low, screen
brightness set to indoor mode) and at least 10 200J discharges (with
one minute of CPR between discharges)
•
At least 30 minutes operating time (at 20 °C± 5 °C of ambient
temperature, wireless function off, voice volume set to low, screen
brightness set to indoor mode) and at least 6 360J discharges (with
one minute of CPR between discharges)
Battery standby life
Standby life
Testing condition
5 years
The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, performing auto test every
day, equipment not in use, not sending selftest report
3 years
The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, performing auto test every
day, equipment not in use, sending selftest report
every week through the wireless network
2 years
The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, performing auto test every
day, equipment not in use, sending selftest report
every day through the wireless network
Waveform storage
Up to 5 hours of ECG waveforms
Events
Up to 500 events
Voice recording
Up to 1 hour
CPR data
Up to 5 hours
Selftest reports
1000 records
Summary of Contents for BeneHeart C Series
Page 2: ......
Page 10: ...4 This page intentionally left blank ...
Page 14: ...1 4 This page intentionally left blank ...
Page 22: ...3 4 This page intentionally left blank ...
Page 34: ...6 4 This page intentionally left blank ...
Page 40: ...8 4 This page intentionally left blank ...
Page 42: ...9 2 9 2 Miscellaneous Description Model PN Disposable battery LM34S002A 022 000411 00 ...
Page 50: ...A 8 This page intentionally left blank ...
Page 64: ...E 4 This page intentionally left blank ...
Page 68: ...F 4 This page intentionally left blank ...
Page 70: ...G 2 This page intentionally left blank ...
Page 72: ...H 2 This page intentionally left blank ...
Page 74: ......
Page 75: ...I 1 I Declaration of Conformity ...
Page 76: ...I 2 ...
Page 77: ......
Page 78: ...P N 046 012619 00 2 0 ...