8 - 3
8.3.2
Auto Test
The equipment with a battery installed carries out auto test at the configured time even when powered off to
check the equipment’s operational performance and alert the operator if a problem exists. Auto test is initiated
at 3:00 am every day by default.
The equipment provides no voice prompts during auto test. The test result can be checked on the status
indicator:
■
Flashes in green: the auto test passes, an auto-test report is saved automatically when the test is
completed.
■
Flashes in red: the auto test fails. If the equipment is connected to the AED ALERT system, an auto-test
report is saved and uploaded automatically to the system when the test is completed.
Mindray recommends you check the status indicator every day, and record the result according to
CAUTION
•
With the equipment powered off, auto test can be performed only when the battery is installed.
NOTE
•
When the equipment is placed at a temperature below -20°C, the auto test cannot be carried out,
and incorrect status indication could result.
8.3.3
Electrode Pads Check
The expiration date of electrode pads should be checked every month. You can check the expiration date from
the pad expiration window, and record it according to
.
8.4
Disposing the Equipment
Dispose of the equipment and its accessories when its service life is reached. Follow local regulations regarding
the disposal of such product.
WARNING
•
For disposal of parts and accessories, where not otherwise specified, follow local regulations
regarding disposal of hospital waste.
Summary of Contents for BeneHeart C Series
Page 2: ......
Page 10: ...4 This page intentionally left blank ...
Page 14: ...1 4 This page intentionally left blank ...
Page 22: ...3 4 This page intentionally left blank ...
Page 34: ...6 4 This page intentionally left blank ...
Page 40: ...8 4 This page intentionally left blank ...
Page 42: ...9 2 9 2 Miscellaneous Description Model PN Disposable battery LM34S002A 022 000411 00 ...
Page 50: ...A 8 This page intentionally left blank ...
Page 64: ...E 4 This page intentionally left blank ...
Page 68: ...F 4 This page intentionally left blank ...
Page 70: ...G 2 This page intentionally left blank ...
Page 72: ...H 2 This page intentionally left blank ...
Page 74: ......
Page 75: ...I 1 I Declaration of Conformity ...
Page 76: ...I 2 ...
Page 77: ......
Page 78: ...P N 046 012619 00 2 0 ...