A - 6
360 J defibrillation waveform into impedance of 25Ω, 50Ω, 75Ω, 100Ω, 125Ω, 150Ω, 175Ω
Selected energy accuracy
Impedance
Energy
25Ω
50Ω
75Ω
100Ω
125Ω
150Ω
175Ω
Accuracy
10 J
9.7 J
10 J
9.7 J
9.3 J
8.9 J
8.5 J
8.1 J
±10% or ±2J,
whichever is
greater
15 J
15 J
15 J
15 J
14 J
13 J
13 J
12 J
20 J
20 J
20 J
20 J
19 J
18 J
17 J
16 J
30 J
29 J
30 J
29 J
28 J
27 J
25 J
24 J
50 J
49 J
50 J
49 J
47 J
45 J
43 J
41 J
70 J
68 J
70 J
68 J
65 J
62 J
60 J
57 J
100 J
97 J
100 J
97 J
93 J
89 J
85 J
81 J
150 J
146 J
150 J
146 J
140 J
134 J
128 J
122 J
170 J
166 J
170 J
166 J
159 J
151 J
145 J
138 J
200 J
195 J
200 J
195 J
187 J
178 J
170 J
163 J
300 J
292 J
300 J
292 J
280 J
267 J
255 J
244 J
360 J
351 J
360 J
350 J
336 J
321 J
306 J
293 J
Charge time (at 20 °C± 5 °C of ambient temperature)
Battery status
From lid opened to
charge done
From initiation of rhythm
analysis to charge done
From initial power
on to charge done
200J
360J
200J
360J
200J
New battery
<8 s
<15 s
<5 s
<12 s
<7 s
New battery after 15 times of
360J discharges
<8 s
<15 s
<5 s
<12 s
<7 s
Time (ms)
V
o
lt
age (
V
)
Summary of Contents for BeneHeart C Series
Page 2: ......
Page 10: ...4 This page intentionally left blank ...
Page 14: ...1 4 This page intentionally left blank ...
Page 22: ...3 4 This page intentionally left blank ...
Page 34: ...6 4 This page intentionally left blank ...
Page 40: ...8 4 This page intentionally left blank ...
Page 42: ...9 2 9 2 Miscellaneous Description Model PN Disposable battery LM34S002A 022 000411 00 ...
Page 50: ...A 8 This page intentionally left blank ...
Page 64: ...E 4 This page intentionally left blank ...
Page 68: ...F 4 This page intentionally left blank ...
Page 70: ...G 2 This page intentionally left blank ...
Page 72: ...H 2 This page intentionally left blank ...
Page 74: ......
Page 75: ...I 1 I Declaration of Conformity ...
Page 76: ...I 2 ...
Page 77: ......
Page 78: ...P N 046 012619 00 2 0 ...