6
•
Do not attempt to repair the pump. Service should be performed by trained and authorized
personnel only.
•
Place the pump on a clean, flat, stable, horizontal surface, away from any source of excessive
heat and close to an easily accessible, properly grounded power supply outlet.
•
Do not expose the pump or the communication hub to liquid. If this happens, immediately
switch off and disconnect the pump from the power outlet, and then decontaminate the
pump and communication hub surfaces. For cleaning instructions, see the
Steritest Symbio
Pumps User Guide.
•
Never expose the equipment to extreme temperatures. Operating temperature must be
between 15 °C and 40 °C (59 and 104 °F).
•
Use only accessories and replacement parts designed for the pump. See
in this user guide. For additional accessories and replacement parts, see
the
Steritest Symbio Pumps User Guide. Using accessories not designed for the pump could
result in user injury or damage to the instrument.
•
The Steritest
™
Symbio Pump has been designed to be used with the Steritest
™
devices and
accessories.
•
When filtrating hazardous liquids, wear and use proper protective clothing and equipment for
the handling and the disposal of the liquid to be filtrated.
•
In case of skin contact with the filtrated liquid, refer to the safety datasheet of the filtrated
liquid for first aid measures.
•
Dispose the filtrated liquids according to local regulations.
•
Do not use the Steritest
™
Symbio Pump to filtrate flammable products.
•
Never touch the display or control panel with a sharp object.
•
Before cleaning, shut down the Steritest
™
Symbio Pump and switch off and disconnect the
communication hub from the power source.
•
The power supply must be protected by a fuse below the main connection.
•
The electrical installation must comply with local standards.
•
Use an electrical surge protector to prevent damage to the system.