Connect cable 6 to the electronic board.
This ribbon cable is permanently attached to the pump upper part. It goes to the right of
the stator motor reducer block, goes through the cable holder located at the right of the
electronic board, and is connected to the electronic board.
Câble 6
45.
Place the rotor motor reducer block in its housing by aligning the coupling pieces located
on the rotor motor reducer block and at the bottom of the pump upper part.
NOTE
Do not turn the coupling piece located at the bottom of the pump upper part. If the
position of the rotor motor reducer belt has not been modified during the disassembly
step, the alignment should be okay. If necessary, turn the rotor motor reducer belt to
adjust the orientation of the coupling piece located on the rotor motor reducer block.
Coupling pieces