![Millipore Steritest Symbio ISL Installation Manual Download Page 8](http://html1.mh-extra.com/html/millipore/steritest-symbio-isl/steritest-symbio-isl_installation-manual_1796934008.webp)
8
Materials of
construction
Bottle holder
Bottle holder support rod
316L Stainless steel
Bottle holder basket
316L Stainless steel
Bottle holder fastening screw clip
Polyphenylsulfone (PPSU)
Communication hub
Housing
316L Stainless steel with epoxy paint
Feet
Nylon and ethylene vinyl acetate (EVA)
USB port
Nickel-plated brass and polybutylene
terephthalate (PBT)
Fuse holder
Thermotin-plated copper alloy
Ethernet port
Nickel-plated brass and polybutylene
terephthalate (PBT)
Communication ports (for
footswitch, etc...)
Nickel-plated brass and polybutylene
terephthalate (PBT)
Power inlet
Polyamide 6.6 and nickel-plated steel
Pump connection cable
Polyvinyl chloride (PVC)
Pump connection cable grommet
Nickel-plated brass
Pump connection cable connector
Chrome-plated brass
Electrical
Specifications
Power supply voltage
100 to 240 Volt AC, 50/60 Hz
Input to pump
24 Volt DC
Power
140 W
Operating
Requirements
Ambient temperature
15 to 40 °C (59 to 104 °F)
Relative humidity
< 90%
Altitude
< 2000 meters (6561 feet)
Filtration units
Compatible with all Steritest
™
,
Sterisolutest
®
, Steridilutor
®
and liquid
transfer kits
Protection type (IEC 60529-2004)
IP64 for the pump
Sound intensity
<70 dB