16.
Gently pull this cable holder and release the cables.
17.
Disconnect cables 1, 2, and 8 from the main electronic board.
18.
Locate the 4 screws on the stator motor reducer block that secure it to the pump (see
illustration above).
19.
Use a #3 hex key to unscrew and remove the 4 screws and washers.
20.
Remove the stator motor reducer block from the pump.
1
2
8
Stator motor
reducer block