Release 08/2014 - Spectroquant
®
Multy
Pipette 10 ml of the
sample into a 24-mm
cell.
Add 1 level blue micro-
spoon of Br
2
-1, close
with the screw cap.
Measuring range: 0.10 – 5.00 mg/l Br
2
24-mm
cell
Bromine
100605
Test
Shake the cell vigor-
ously to dissolve the
solid substance.
Fill approx. 10 ml of
distilled water into a
24-mm cell (do not add
any reagents!), close
with the screw cap.
(Blank cell)
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Reaction time:
1 minute
Press Enter to start
the countdown.
Check the pH of the
sample, specified range:
pH 4 – 8.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Select method 9 0 .
Important:
Very high bromine concentrations in the sample produce
yellow-colored solutions (measurement solution should be
red) and false-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) a freshly prepared standard
solution
can be used (see section “Standard solutions”).
Insert the blank cell into
the cell compartment.
Align the mark on the
cell with that on the pho-
tometer. Press Zero .