Release 08/2014 - Spectroquant
®
Multy
Place 1 level blue
microspoon each of
NO
3
-1 into two dry
16-mm cells.
Nitrate
114773
Test
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) we recommended to use
Spectroquant
®
CombiCheck 10 and 20, Cat.No. 114676
and 114675, or the Standard solution for photometric
applications, CRM, Cat.No. 125036 and 125037.
Ready-for-use nitrate standard solution Certipur
®
,
Cat.No. 119811, concentration 1000 mg/l NO
3
–
, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
Measuring range: 0.5 – 15.0 mg/l NO
3
-N 16-mm
cell
2.2 – 66.4 mg/l NO
3
16-mm
cell
Add to each cell 5.0 ml
of NO
3
-2 with pipette,
close with the screw cap
Select method 3 2 1 .
Insert the cell containing
the sample into the cell
compartment. Press
Test .
Reaction time:
10 minutes
Press Enter to start
the countdown.
Shake both cells vigor-
ously for 1 minute to
dissolve the solid sub-
stance.
Add to one cell very
slowly 1.5 ml of the
sample with pipette,
close with the screw
cap, and mix briefly.
Caution, cell becomes
very hot !
Add to the second cell
very slowly 1.5 ml of
distilled water with pi-
pette, close with the
screw cap, and mix
briefly.
Caution, cell becomes
very hot !
(Blank cell)
Insert the blank cell into
the cell compartment.
Press Zero .