Release 08/2014 - Spectroquant
®
Multy
Measuring range: pH 6.4 – 8.8
16-mm cell
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) buffer solution pH 7.00
Certipur
®
, Cat.No. 109407, can be used.
Pipette 10 ml of the
sample into a round
cell.
Add 4 drops of pH-1,
close with the screw
cap, and mix. Attention !
The reagent bottle must
be held vertically by all
means !
101744
Cell Test
pH
Fill approx. 10 ml of
distilled water into an
empty 16-mm cell (do
not add any reagents!),
close with the screw
cap.
(Blank cell)
Select method 3 6 0 .
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Insert the blank cell into
the cell compartment.
Press Zero .