Chapter 5.3 | Page 1 of 6
5.3
Printing measurement results
5.3.1 Setting the print parameters
The Spectroquant® Multy Colorimeter can print out data on a
printer with a serial interface via the RS232 port.
The standard settings of the printer type used should be checked
before printing out data. The usual settings are as follows:
Data bits:
8
Parity:
none
Baud rate:
dependent on the printer type
e.g.
LQ 300 matrix printer: 4800
DP 1012 ticket printer: 19200
The printing parameters of the Spectroquant® Multy
Colorimeter must be aligned to match these settings
accordingly. This is done in the following manner:
Press the keys
[Mode] [2] [9]
in succession.
Confirm your selection by pressing
[Enter]
.
The display now shows:
Press the
[1]
key to set the flow control.
The display now shows:
Press the
[
]
or
[
]
arrow keys to select the desired
settings depending on the type of printer in question
(Xon/Xoff, none, hardware). The setting “none” must be
selected for the LQ 300 unit.
Confirm by pressing the
[Enter]
key.
Press
[Esc]
to exit this mode.
< printing params.>
1:Flow control
2:Baud rate
cancel:ESC
< Flow control>
is:Xon/Xoff
select:
save:
cancel:ESC