Release 08/2014 - Spectroquant
®
Multy
Add 5 drops each of
Zn-1K into two reaction
cells, close with the
screw cap, and mix.
114566
Cell Test
Measuring range: 0.20 – 5.00 mg/l Zn
16-mm cell
Zinc
Add to one cell 0.50 ml
of the sample with pi-
pette, close with the
screw cap, and mix.
Add to each cell 5 drops
of Zn-2K, close with the
screw cap, and mix.
Add to the second cell
0.50 ml of distilled water
with pipette, close with
the screw cap, and mix.
(Blank cell)
Select method 5 4 1 .
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Reaction time:
15 minutes
Press Enter to start
the countdown.
Check the pH of the
sample, specified range:
pH 3 – 10.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Important:
For the determination of total zinc a pretreatment
with Crack Set 10C, Cat.No. 114688, or Crack Set 10,
Cat.No. 114687, and thermoreactor is necessary.
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) we recommended to use
Spectroquant
®
CombiCheck 40, Cat.No. 114692.
Ready-for-use zinc standard solution
Certipur
®
, Cat.No.
119806, concentration 1000 mg/l Zn, can also be used
after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 40) is highly recommended.
Insert the blank cell into
the cell compartment.
Align the mark on the
cell with that on the pho-
tometer. Press Zero .