![Millipore Spectroquant Multy Operating Manual Download Page 146](http://html1.mh-extra.com/html/millipore/spectroquant-multy/spectroquant-multy_operating-manual_1796903146.webp)
Release 08/2014 - Spectroquant
®
Multy
114547
Cell Test
Nitrite
Measuring range: 10 – 700 µg/l NO
2
-N 16-mm
cell
33 – 2299 µg/l NO
2
16-mm cell
Pipette 5.0 ml of the
sample into a reaction
cell, close with the
screw cap.
Shake the cell vigor-
ously to dissolve the
solid substance.
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) ready-for-use nitrite stan-
dard solution Certipur
®
, Cat.No. 119899, concentration
1000 mg/l NO
2
–
, can be used after diluting accordingly as
well as the Standard solution for photometric applications,
CRM, Cat.No. 125041.
Fill approx. 10 ml of
distilled water into an
empty 16-mm cell (do
not add any reagents!),
close with the screw
cap.
(Blank cell)
Select method 3 3 0 .
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Reaction time:
10 minutes
Press Enter to start
the countdown.
Insert the blank cell into
the cell compartment.
Press Zero .
Check the pH of the
sample, specified range:
pH 2 – 10.
If required, add dilute
sulfuric acid drop by
drop to adjust the pH.