- 31 -
Registering UV Lamp timer
Introduction
The timer used for the UV 185 nm Lamp must be reset when the System is
installed.
If this is not done, then the message indicating that a Lamp replacement is
needed is shown too early.
NOTE:
Before doing this, make sure that the date and time have been checked for
accuracy.
Procedure
This procedure shows how to reset the timer used for the UV 185 nm Lamp.
Step Action
Diagram
1
Place the System in STANDBY
Mode.
2 Select
Menu.
Press
.
3 Select
Maintenance.
Press
.
4
Select Install UV 185 nm
Lamp.
Press
.
Continued on next page