- 33 -
Calibrating the Flowrate
Introduction
The Milli-Q Water flowrate should be calibrated when the System is installed.
A 1 Litre graduated cylinder is needed.
Procedure
Follow the steps below to perform a Flow Calibration.
Step Action
Diagram
1
Go to STANDBY Mode.
2 Select
Menu.
Press
.
3
Enter the Manager Menu.
See the
Software Chapter
to
learn how to enter the Manager
Menu.
4 Select
Setup.
Press
.
5
Select Flow Calibration.
Press
.
Continued on next page