- 80 -
Summary of Alarm messages,
Continued
Alarm
messages
(continued)
LCD message
What it means
MILLI-Q RES > MAX
The Milli-Q Water resistivity is out
of measurement range.
Contact Millipore.
MILLI-Q T < MIN
The Milli-Q Water temperature is
out of measurement range.
Contact Millipore.
MILLI-Q T > MAX
The Milli-Q Water temperature is
out of measurement range.
Contact Millipore.
POD LOCKED
The POD Unit microswitch is
locked.
Push the Plunger all the way down
and release.
Q-GARD PACK OUT
The Q-Gard Pack is not installed
correctly or it has been removed.
The System stops operating.
Verify that the Q-Gard Pack is
installed correctly.
Contact Millipore if the problem
continues.
QUANTUM CARTRIDGE OUT
The Quantum Cartridge is not
installed correctly or it has been
removed. The System stops
operating.
Verify that the Quantum Cartridge
is installed correctly.
Contact Millipore if the problem
continues.
TANK EMPTY
The System has detected an empty
Reservoir.
Refill the Reservoir.
Verify that the Reservoir level
sensor is plugged into the System
Cabinet.
WATER DETECTED
A Water Sensor (an accessory
connected to the System) has
detected water. The System stops
operating.
Clean up the spilled water.
Make sure the source of the leak is
fixed.