OPERATION
REV.0.01_COD.650184_A4
Page 25
WARNING
Hold the door during opening and closing. Jerking the door open or
slamming it shut may break the glass.
Insert baskets into the cabinet slowly to avoid breaking the glass.
Before opening the door, check that the door area is free of obstacles.
Emergency release
The emergency release may only be used when it is no longer possible to open the door normally,
e.g., in the event of a power failure.
WARNING
If the emergency release is operated during a program sequence, hot water
and process chemicals can escape.
The wash items, the load carriers, and the wash cabinet may be very hot.
Danger of scalding, burning, and chemical burns. Where disinfecting agents
are used, there is also a danger of inhaling toxic fumes.
In the event of a power failure or if the door lock is damaged, the door can be opened as follows:
1.
An emergency release device is located between the door and the
cover plate and is identified by a sticker (see image).
2.
Insert the tool contained in the accessory kit into the emergency
release opening.
3.
Prize the tool to the left until you hear the door open.
4.
To close the door, continue to press in the tool as described in point
3.