Additional settings
87
If you select
Main wash
or
Final rinse
, the following settings can be
altered:
Reset
Temperature
Holding time
–
Temperature
/
Final rinse temperature
The temperature of the selected block can be altered.
–
Holding time
The holding time of the selected block can be altered.
Select an option using the
and
arrow buttons and confirm your
selection with
OK
.
Holding time
Temperature
50
(30 - 70
°C
°C)
)
Min
(1 - 10
3
Min
The setting value is entered in increments of 1. The possible range is
shown in the bottom line of the display.
Dispensing of process chemicals occurs at a default dispensing
temperature set at the factory. If process chemicals are to be
dispensed in this wash block, the lowest temperature that can be set
will be the dispensing temperature. It is not possible to set a lower
value.
Use the arrow buttons
(higher) and
(lower) to set the value.
Press
OK
to save the setting.