Installation
120
Electromagnetic compatibility (EMC)
The machine has been tested for electromagnetic compatibility (EMC)
in accordance with EN 61326-1 and is suitable for operation in
commercial environments, such as laboratories and other similar
environments which are connected to the mains power supply.
The machine's high frequency (HF) energy emissions are very low and
are therefore unlikely to interfere with other electronic appliances in
the vicinity.
Flooring in the installation area must be wood, concrete or tiled.
Synthetic flooring must be able to withstand a relative humidity level
of 30 % to minimize the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or laboratory / hospital environment and should
deviate from the nominal voltage by a maximum of +/- 10 %.