Chapter 15 Troubleshooting
157
automatically standby at the set time.
13.
Oscilloscope will automatically shut down after no operation for a period of time:
⚫
Check whether the shutdown time setting is Disabled; if it is not Disabled, it will automatically shut
down when the set time is reached.
14.
CSV files cannot be selected when loading reference:
⚫
CSV files are not the supported format that can be loaded into reference channels.
15.
Tap the button during the use of oscilloscope, there is no beep sound:
⚫
Check whether the sound volume setting is correct.
16.
Oscilloscope backlight has low brightness:
⚫
Check whether the backlight settings are correct.
17.
A waveform being moved changes abruptly:
⚫
Check whether the picture is displayed in full screen.
18.
Turn off the channel at Auto state:
⚫
This phenomenon is normal. At Auto state, the channel with an amplitude less than 10mV will be turned
off.
19.
Function buttons are pressed without response:
⚫
Check whether the picture is displayed in full screen.