1.4 Specifications
Electrical
AC Input:
90-132 VAC, 47-63 Hz, single phase
200-265 VAC, 47-63 Hz, single phase
Line Cord:
IEC320/UL498 input connector. Replacement cord must have safety-ground wire, and must
have a current rating greater than the fuse ratings listed below.
Power Consumption:
25 VA maximum
Fuse Ratings:
115 VAC 2 @ 1.0 A
230 VAC 2 @ 0.5 A
Fuse:
Installed in AC input module fuse tray in rear panel
5 x 20mm type T cartridge fuses, 250V
Computer Interface:
RS232 using proprietary communication cable, (see section 5.4)
Recirculator Interface (optional):
Standard, molded, 25 pin, male-female, straight through interface cable
Ultrasonic Power Output (optional):
Coaxial cable, BNC connectors
Lasers
Wavelength:
780 nm
Power:
3 mW nominal
Environmental
Temperature:
10 to 35ºC
Humidity:
90% RH, non-condensing maximum
Storage Temperature:
–10 to 50ºC (dry only)
Pollution:
Degree 2
Connection Ports, Wet Operation
Fittings:
3/8" 316 stainless steel
Connection Ports, Dry Operation
Vibratory Inlet:
Open funnel
Aspiration Inlet:
¼” opening (base of funnel), friction-fit tube fitting
Vacuum Connection:
Proprietary quick-disconnect fitting, 1¼” tube (standard vacuum hose)
If supplied by user, vacuum source must exceed 50 CFM, 50 “H
2
O
Eductor Air Connections:
4mm push-to-connect fittings
Eductor Air Supply:
100 psi (689 kPa) maximum pressure
5 CFM at 50 psi (345 kPa) minimum flow rate
Free of dry contaminants, moisture and oil
Particle Size
Particle Size:
3000
µ
m maximum dimension of spherical or non-spherical particles