3 Command Prompts
Option
Description
Range:
1.0e-40
to
9.99999e-37 cm
3
Default:
4.25146e-38 cm
4
Enter the
adsorbent
spherical parameters.
Range:
1.0e-40
to
9.99999e-37 cm
3
Default:
6.05156e-38 cm
3
For more details on the spherical parameters, refer to the Calculations
document found on the Micromeritics web page
(
).
Cheng-Yang corr.
Choices:
Yes, No
If
Yes
is selected, the Cheng/Yang correction is applied to the pore size
analysis. This correction substitutes the Langmuir equation of state for
Henry’s Law in the Horvath- Kawazoe derivation.
Report Options (continued)
3 - 32
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021