2 Operational Procedures
A
NALYSIS
R
ESULTS
Press
CHOICE
to display the options. Press
ENTER
to select.
There are several methods for working with analysis results:
n
View
n
n
Transmit
n
V
IEW
A
NALYSIS
R
ESULTS
As the analysis is performed, operational status messages are displayed. An example message is
shown below:
A. Number of points collected
B. Number of points requested
C. Current pressure (mmHg)
D. Quantity adsorbed (cm
3
STP)
E. Rate of change for quantity adsorbed
During analysis, the quantity (Q) value equals the quantity adsorbed by the sample in cm
3
STP. At
all other times, the Q value equals the reading of the differential transducer in cm
3
STP. This
reading indicates the amount of gas in the balance side of the system above the amount in the
sample side of the system. This value is usually near zero (0) when an analysis is not in progress.
When the analysis is complete, the analyzer beeps three times. If data have been collected and
surface area or pore volume calculations have been selected, press
CHOICE
to view the results.