Control Commands
optoNCDT ILR 1191
10.3.14 QA – Analog Output
Format: QAx y [Enter]
QA parameterizes the behavior of the QA analog output, see Chap.
5.3.2
.
i
The parameter SF affects the analog output.
Available for setting are the lower limit x and the upper limit y of a current range from 4 to 20 mA. The scale
factor SF, see Chap.
10.3.5
, does not affect the limits.
The lower limit may both be smaller and greater than the upper limit, accordingly, the current range will
reverse, see Chap.
5.3.2
.
Inputs of identical limits will be ignored and not accepted.
Request
QA
Set
QAx y
Parameter value
range x:
float32; resolution: 0.001
Parameter value
range y:
float32; resolution: 0.001
Standard
1.000 300.000
The sensor performs no check for plausibility of QA settings.
10.3.15 BR – Baud Rate
Format: BRx [Enter]
The command facilitates conversion to other serial baud rate x.
Following a change in the baud rate, a cold start is not necessarily required.
Request
BR
Set
BRx
Parameter value
range x:
9600, 19200, 38400, 57600, 115200, 230400 oder 460800
Standard
115200